精确可编程序的注射器微量注射泵 
(USA, Era Pump Systems) 特价   | 
注射器微量注射泵可用于动物实验,对动物进行微量注射,可控制流速, 可通过程序控制注射流量。

主要性能特征:
1. 单机操作也可电脑控制 (RS-232插头)
2. 注射也可回抽 (双向)
3. 用1~60cc注射器
4. 可选择注射流量及流速控制程序,有41种选择
5. 多个泵连接组合成一个系统 (双筒), 最大可达100泵;
6. 单筒NE-1000、6筒NE-1600、8筒NE-1800整机选择.
7. 注射速率pumping rate: 
NE-1000: 0.73 ml/小时~2100ml/小时 
NE-1600: 0.568 ml/小时~1337ml/小时 
NE-1800: 0.568 ml/小时~380 ml/小时 
规格表Descriptions 
- Model : NE-1000 NE-1600 NE-1800 
 - Syringe sizes: 1 ~ 60 cc    1 ~ 60 cc 1 ~ 10 cc 
 - Number of syringes: 1~ 6 8 
 - Electrical 
 - Motor type: Step motor 
 - Motor steps per revolution: 200 
 - Motor to drive screw ratio: 1/4 
 - Drive screw pitch: 24 revolutions 
 - Advance per motor step: 1.323 um 
 - DC connector: 2.5 mm, center positive 
 - Voltage at DC connector: 12V DC at full load 
 - Amperage: 1000 mA at full load 
 - Power supply output rating: 12V DC @ 1000 mA 
 - Dimensions: 
 - NE-1000: 22.86 cm x 14.605 cm x 11.43 cm 
 - NE-1600,1800: 26.35 cm x 38.42 cm x 13.34 cm 
 - Weight: NE-1000 (1.63 kg) NE-1600,1800 (4.6 kg) 
 - Operational 
 - Maximum speed: 3.2226 cm/min 
 - Minimum speed: 0.0262 cm/hr 
 - pumping rate: 
 
NE-1000: 0.73 ml/hr~2100ml/hr 
NE-1600: 0.568 ml/hr~1337ml/hr 
NE-1800: 0.568 ml/hr~380 ml/hr 
- Maximum force: 160 lbs at minimum speed 
 
30 lbs. at maximum speed 
- Number of Program Phases: 41 
 - RS-232 pump network: 100 pumps maximum 
 - Syringe inside diameter range: 0.100 ~ 50.00 mm 
 
| 
 | 
 | 
- Infuses and withdraws 
 - Selectable pump rate units: µl/hr, µl/min, ml/hr, ml/min. 
 - Change pumping rate and direction while pumping. 
 - Dispense a specified volume. 
 - Syringe purge mode will infuse or refill syringe at top speed to purge air from the line or to fill the syringe. 
 - Audible buzzer can be programmed to alert when an alarm condition occurs or the pumping program completes. 
 - Power failure mode restarts a pumping program interrupted by a power failure. 
 - Stall detection stops pump when the motor's operation is impeded (not on NE-500 model) 
 - Non-volatile memory restores all setup parameters and the pumping program, on power up. 
 - Elastomer raised keypad with optional beeping whenever the keypad is pressed. 
 - CE certified 
   | 
| 
 | 
 | 
- Pre-program a set of production dispensing protocols. The operator then selects the required dispense protocol 
 - Keypad Lockout: Prevent unauthorized or accidental changes to the settings by requiring a "Lockout Disable" key to be inserted to change settings. 
   | 
| 
 | 
 | 
- 41 programmable phases 
 - Enter programs from the keypad or from a computer. 
 - Download or upload pumping programs from or to a computer. Store dispensing programs in a file on your computer, then download them to one or more pumps. Specify in the file to send a different pumping program to each pump in the pump network. 
 - Pre-program dispense volumes. Each volume can be different. 
 - Automatically change pumping rates or pumping direction. 
 - Ramp up or down the pumping rate. 
 - Set timed delays between dispenses. 
 - Automatically pause the program and wait for the user to continue the dispense. 
 - Synchronize dispenses with other equipment or pumps to change pumping rates in reaction to a sensor or signals from other pumps. 
 - Send logic signals to other equipment. Change pumping rates in reaction to a sensor. 
 - Program the audible alarm to beep at any time to alert the operator. 
   | 
| 
 | 
| 
 | 
 | 
| 
 | 
- Network up to 100 pumps from a single communications port on a computer. Baud rates up to 19200. 
 - Free terminal emulator program allows communications with syringe pump. 
 - Directly control one or more pumps from a computer, including setting the pumping rate, direction and volume. Set the operational and setup configurations. Also the TTL interface can be queried and the programmable output set. 
 - Using a computer, each pump in the pump network can be individually controlled and monitored. Or, the computer can download a new pumping program to each pump which would automatically reconfigure the dispensing system for a new job. 
 - Downloading control programs to the pump would unload computer overhead to the pump. Computer resources could then be used elsewhere while the pumps operate independently. 
 - 2 modes of communications: Basic and Safe 
- Basic: Basic master-slave interface. Easy to implement. Can communicate from any terminal emulator. 
 - Safe: A reliable interface for a real-world environment. 
- - High quality interface which detects communication errors via a 16 bit CCITT CRC. 
 - - The pump will automatically transmit an alarm packet when a pump alarm occurs. For instance, if the pump stalls or resets, a message is immediately sent to the computer. 
 - - A configurable communications time-out will stop the pump and alarm if the communications link is broken. 
 - - Packet inter-character time-out to recover packet synchronization after communication glitches. 
    
 - Software Developer friendly communications protocol. 
- Consistent and well defined communications packets. 
 - Constant start and end characters. 
 - Packets structured to be easily parsed 
     | 
| 
 | 
| 
 | 
 | 
| 
 | 
- Control the pump from a foot switch or timer. 
 - Output signals can be used to switch a valve from an output to an input reservoir. 
 - Keypad Lockout: Prevent unauthorized or accidental changes to the settings by requiring a "Lockout Disable" key to be inserted to change settings. 
 - Pumping direction control 
 - Programmable output signal 
 - Programmable responses to input signals 
 - Configurable start and stop trigger: 
- Start or stop the pump on each contact closure to ground. For example, a foot switch. 
 - Start the pump on closure to ground and stop the pump on opening the connection. For example a timer relay that starts the pump whenever closed and stops the pump whenever open. 
 - Only start the pump on each contact closure to ground. For example, a laboratory animal trained to press a lever can receive a drug dosage from the pump and the pump will not stop or deliver another dosage until it is permitted by the pumping program entered by the user 
  
 - Configurable trigger input: 
- Edge trigger: Starts or stops the pump, such as from a foot switch 
 - Level trigger: Start or stops the pump, such as from a timer or relay control 
 - Start only trigger: Edge trigger will only start the pump, such as from a lab animal trained to press a lever to receive a dosage. 
    
 | 
 | 
 | 
 | 
Special Application: 双筒连续灌注Continuous Infusion/Dual Syringe Pump
注射器加热装置
SYRINGE-HEATER-KIT | 
 注射器加热装置  | 
Description:
- Heating device for syringes that require temperature-controlled dispensing 
 - Heats syringe to over 100 degrees Celsius 
 - Digitally set the heater setpoint 
 - Temperature sensor in heating pad 
 - Setpoint retained in memory 
 - Controller will heat the syringe and hold the set temperature using an on/off or PID control algorithm 
 - Multiple syringe heater pads can be attached to the control unit 
 - Heated syringe can be mounted on a syringe pump 
 - Syringe Heater Kit includes Control-Unit power cable and Primary-Syringe-Heater-Pad 
   |